CeBr3 (cerium bromide) is an inorganic scintillator for gamma-ray spectroscopy. It gives energy resolution close to LaBr3(Ce) with much lower intrinsic background, because the crystal contains no lanthanum and therefore no La-138. ATR Crystal can supply CeBr3 crystals on request through its partner crystal growers; size, encapsulation and test scope are confirmed in each quotation.
This guide summarises published CeBr3 properties, compares CeBr3 with LaBr3(Ce) and NaI(Tl), and lists what to specify when requesting a crystal.
CeBr3 scintillator properties
The values below are published reference values (Berkeley Nucleonics and SCIONIX data sheets). They are not ATR Crystal batch test results; light output, decay time and resolution all vary with crystal size, encapsulation and readout.
| Property | Reference value |
|---|---|
| Chemical formula | CeBr3 (cerium(III) bromide), intrinsic scintillator |
| Density | About 5.2 g/cm3 |
| Light output | About 60,000 photons/MeV |
| Decay time | About 18 to 25 ns, depending on crystal size |
| Peak emission | About 370 nm |
| Refractive index | About 2.09 at 380 nm |
| Energy resolution | About 4% FWHM at 662 keV (typical) |
| Hygroscopic | Yes; used in a hermetically sealed housing |
CeBr3 vs LaBr3(Ce) vs NaI(Tl)
| Property | CeBr3 | LaBr3(Ce) | NaI(Tl) |
|---|---|---|---|
| Density (g/cm3) | 5.23 | 5.07 | 3.67 |
| Relative light yield (NaI(Tl) = 100) | 130 | 150 | 100 |
| Decay time | 18 to 25 ns | 16 to 20 ns | About 230 ns |
| Peak emission | 370 nm | 370 nm | 415 nm |
| Energy resolution at 662 keV | About 4% | About 3% | About 7% |
| Intrinsic background | Low; trace Ac-227 contamination | La-138 self-activity plus trace Ac-227 | No intrinsic activity in the crystal itself |
| Hygroscopic | Yes | Yes | Yes |
Source: Berkeley Nucleonics, Choosing a scintillator (Table 1). Treat the figures as typical values for comparison.
When CeBr3 is the better choice
LaBr3(Ce) gives the best resolution of the three, but its own La-138 activity adds counts to every spectrum. For low-activity samples, long counting times or low-background setups, that self-activity can matter more than the extra resolution. CeBr3 removes the La-138 contribution. Its remaining trace Ac-227 background appears mainly as peaks between about 1.5 and 2.2 MeV; Berkeley Nucleonics quotes about 0.002 counts/s/cm3 measured inside a 100 mm lead shield.
Compared with NaI(Tl), CeBr3 offers clearly better energy resolution at 662 keV (about 4% versus 7%) and a decay time about ten times shorter, which helps peak separation and high count rates. NaI(Tl) remains the economical choice for large volumes where about 7% resolution is enough. For a deeper NaI comparison, see LaBr3(Ce) vs NaI(Tl) for gamma spectroscopy and scintillator energy resolution at 662 keV.
Low intrinsic background is not zero background
Trace radioactive contamination is a separate question from the host material. The JRC study of CeBr3 radiopurity detected Ac-227 and other radionuclides in the investigated crystal, including a small La-138 contribution. This is sample-specific evidence, not a specification for every CeBr3 crystal.
A measured detector spectrum also includes environmental radiation, housing and photosensor contributions. The Quarati et al. study of CeBr3 gamma-ray spectrometers compared sample backgrounds using shielding and defined acquisition conditions, which is why the measurement setup must be checked alongside the material name.
What to request in a CeBr3 background report
| Report item | Information needed for comparison |
|---|---|
| Sample identity | Crystal dimensions, batch or sample reference, and whether the measurement used a bare or encapsulated crystal |
| Energy interval | Lower and upper energy limits, thresholds and any regions excluded from the quoted count rate |
| Rate normalisation | Counts/s for the detector, counts/s/cm3, or a rate per energy interval |
| Acquisition | Measurement duration, live-time treatment and reported uncertainty or detection limit |
| Background conditions | Shielding, location, photosensor and housing, with the method used to assess their contributions |
| Data processing | Whether the spectrum is raw or background-subtracted, with the subtraction method identified |
A rate normalised by crystal volume cannot be compared directly with a total detector count rate. Match the units and energy interval first.
Encapsulation and readout
CeBr3 is hygroscopic, so it is used in a sealed housing with an optical window. Review the sealing arrangement, window material, coupling interface and handling procedure before integration, and never expose a bare crystal to room air.
The 370 nm emission suits bialkali photomultiplier tubes. With silicon photomultipliers, check the photon detection efficiency of the chosen SiPM at 370 nm and the transmission of the window; both affect the resolution you will actually measure.
What to send for a CeBr3 quotation
- Crystal size: diameter and length (or a drawing), and quantity.
- Configuration: sealed crystal with window, or crystal coupled to a photosensor; window material and connector requirements.
- Readout: PMT or SiPM type and the electronics you will use.
- Measurement: gamma-ray energies of interest, required resolution and any background acceptance interval.
- Tests: which measurements you need with delivery, so the test scope can be agreed in the quotation.
Send your CeBr3 requirements to ATR Crystal. For other materials in the range, use the scintillation crystal selection guide.
CeBr3 FAQ
Is CeBr3 better than LaBr3(Ce)?
Not in every case. LaBr3(Ce) has slightly better resolution and light output; CeBr3 has much lower intrinsic background. For low-activity or low-background measurements, CeBr3 is usually preferred. For high-activity sources where resolution matters most, LaBr3(Ce) can be the better fit.
What is the energy resolution of CeBr3?
Published data sheets give about 4% FWHM at 662 keV as a typical value. Large crystals and different readouts can give different results.
Is CeBr3 hygroscopic?
Yes. Like NaI(Tl) and LaBr3(Ce), it must be sealed against moisture.
References
- Lutter et al., Radiopurity of a CeBr3 crystal used as scintillation detector (2013), DOI: 10.1016/j.nima.2012.11.174.
- Quarati et al., Scintillation and detection characteristics of high-sensitivity CeBr3 gamma-ray spectrometers (2013), DOI: 10.1016/j.nima.2013.08.005.
- SCIONIX CeBr3 data sheet, revision dated December 3, 2021.
- Berkeley Nucleonics, Cerium bromide (CeBr3) scintillators and Choosing a scintillator.
These publications describe the authors’ measurements and are not ATR Crystal batch inspection reports.