NaI(Tl) Detector Efficiency Calculator

Estimate how many gamma rays from a point source interact in a cylindrical NaI(Tl) crystal. Choose the crystal size, the source distance and the energy (20 keV to 3 MeV).

Crystal size (diameter x height)

Measured on the crystal axis to the crystal itself. Add the housing window and any gap (often a few millimetres).

Gamma-ray energy

Results

Total efficiency-
Geometric efficiency-solid angle / 4π
Intrinsic efficiency-total / geometric

Total efficiency versus energy for this crystal and distance. The dot marks the selected energy.

Total efficiency is the fraction of all emitted photons that interact in the crystal. The full-energy peak efficiency is lower, because some interactions deposit only part of the energy. Bare crystal, isotropic point source on the axis, no housing, window or reflector absorption.

Quote this NaI(Tl) size

How the calculation works

The calculator treats the crystal as a bare NaI(Tl) cylinder (density 3.67 g/cm³) and the source as an isotropic point on the crystal axis. For every direction that reaches the crystal, it finds the path length through the crystal and the probability that the photon interacts along that path, 1 − exp(−μx), then integrates over all those directions.

μ is the total attenuation coefficient of NaI without coherent scattering, from the NIST XCOM database, interpolated on a log-log scale with the iodine K edge at 33.17 keV. Coherent scattering is left out because it does not deposit energy.

Check against published values

Results for a 3″ × 3″ crystal (76.2 × 76.2 mm), compared with the calculated values of Vegors et al. (1958) and Heath (1964) as tabulated by Tarım, Gürler and Yalçın (2018):

Source position Energy This calculator Published
On the crystal face 662 keV 0.364 0.362 (Vegors et al.)
On the crystal face 1332 keV 0.293 0.293 (Vegors et al.)
10 cm from the face 662 keV 0.0202 0.0198 (Heath)
10 cm from the face 1332 keV 0.0164 0.0162 (Heath)

Other published calculations listed in the same paper range from 0.0183 to 0.0207 at 662 keV and 10 cm, and from 0.0162 to 0.0169 at 1332 keV.

What the calculator does not include